T1 measurement by inversion recovery
Inversion Recovery
Here the spins are flipped to the 180° position, and sampled at different times until the spins are back in equilibrium. This will give a characteristic set of spectra, which can be used to accurately measure T1-Times for the sample. The delay τ is varied and the intensities in the spectra are plotted agains τ.
T1 fitting
Using a nonlinear regression algorithm, the data is fitted to the following model:Which can be simplified (when using a normalised plot which has very little noise) to:
For an inversion recovery experiment 1 < p1 < 2 is normally true. In the optimal case, p1 would be 2, but since in reality one does not get a negative signal which is as strong as the maximum positive signal, the parameter p1 compensates for that. The parameter p2 is equal to T1.
When trying to fit multiple T1-Times in a samle, one simply fits a sum of several of the previous model. For two T1-Times, one would use:
Here p1 is used as bevore, but p2 now serves as a parameter determining how much each of the two components contribute to the relaxation behaviour. The parameters p3 and p4 are the two contributing T1-Times.
The Inversion Recovery Sequence and T1 Measurement
An inversion recovery pulse sequence can also be used to record an NMR spectrum. In this sequence, aIt should be noted at this time that the zero crossing of this function occurs for
T1 measurement by saturation recovery
Saturation Recovery
In the Saturation Recovery experiment, the time it takes for the spins to recover from a saturated state is measured. If one has a probehead where the use of gradients is possible, this is quite easy:However without the possibility to use gradients, a set of several pulses in a short timespan (often with decreasing delays between the pulses in this pulsetrain) can be used to saturate the spin system:
T1 relaxation time measurements are usually done with a simple 180 -tau -90, inversion recovery pulse sequence (see figure). Tau is varied from a small value to a large value and a nonlinear regression is carried out to fit the best T1 value.

1. Call up the pulse sequence "t1ir1d" (Bruker) or "s2pul" (Varian).
2. Set p1 and p2 (Bruker) or PW and P1 (Varian) to the 90 degree and 180 degree pulses , respectively.
Set the recycle delay, d1 (Bruker and Varian) to something you believe is much longer than the T1.
3. Set tau to a very small value (3 microseconds for example). Tau is d7
on a Bruker spectrometer or d2 on a Varian spectrometer.
4. Collect a spectrum and phase it such that all peaks are negative (one
scan is often enough for protons). Store the phase correction.
5. Repeat step 3. increasing d7 (Bruker) or d2 (Varian) until the peak
of interest is nulled. If the peak is negative, tau is too short. If it
is positive, tau is too long.
6. The T1 of the peak of interest is the tau value for the null divided by the natural log of 2.
T1 and T2 are DIFFERENT parameters. Depending on what problem is being addressed, it may be necessary to meaure T1 rather than T2. Although they are often equal in solution (under extreme narrowing conditions), it is always true that T1 is greater than or equal to T2.
T1's go through a minimum when the product of the resonance frequency and the correlation time is approximately equal to 1. This minimum can be observed by measuring T1's as a function of temperature. The same minimum does not exist for T2.
Inversion recovery is used most often because it is the simplest. One can use it to measure the T1's for multiple peaks in the spectrum and the magnetization is followed from all the way from -Z to +z for inversion recovery rather than just 0 to +z in the case of saturation recovery.
////////
No comments:
Post a Comment